#3.1 Development of a CTA conditioner for SiC thermal sensors
Commercial
CTA
electronics are designed around conventional hot wires, with
resistances spanning from a few to 100 Ω at most. Therefore, in
order to operate miniature hot wires, where the resistance is
sometimes orders of magnitude higher, in-house systems have to
be created and adapted for each probe. Currently, the only
research group performing dynamic measurements with miniature
hot-wire probes with adapted control electronics is Markus
Hultmark’s team at Princeton University,Who developed the
NSTAP
family of nanoscale hot wires.
using a custom
CTA,
the
PUCTA
built by Watmuff
(1994)26. A replica
of this system, the
MUCTA
is currently in use at Melbourne University for more precise
investigation of the
NSTAP
probes’ performance Samie et
al. (2018)130. As part of
this work, a control system was built to operate the SiC sensors
in
CTA
mode, with specifically selected components to bring sufficient
voltages to properly heat the 25 kΩ resistors.
#3.1.1 Targeted operating points
The semiconductor nature of the heating and sensing elements raises a challenge for hot-wire applications, because the TCR varies with the temperature. Figure 2.33 in the previous chapter shows the evolution of the central wire’s resistance relative to uniform heating, together with the corresponding TCR. There are four distinct resistance regimes over the investigated temperature span, characterized by alternating positive (metal-like) and negative (semiconductor-like) TCR values. Consequently, unlike metallic wires with constant TCR, the mean temperature of a semiconductor wire under non-uniform heating—typically Joule heating—cannot be inferred from its mean resistance.
#3.1.2 Design of the circuit
Due to larger resistance values, SiC wires have to undergo
large voltage drops to be sufficiently heated,For SiC calorimeters presented above, a 38 mW
heating power generates a ~35 V potential on the wires.
meaning that each active component should
properly function at these voltages. This forbids the use of
commercial
CTA
circuits, which are designed to work with ~1-20 Ω
resistances.Typically 4 Ω for hot wire, 15 Ω for hot-film
probes.
A Perry-type
CTA
control circuit has been designed to drive high-value SiC
resistances, with components able to work up to a 35 V working
point, as shown in figure 3.1 (a).
This enables the bridge to function at a high enough voltage to
properly power the SiC heater. The bridge voltage is controlled
by a 2N2222A
BJT
driven by the controller output. The main differential amplifier
is an ADA4522 whose output is amplified by a
variable gain INA128 instrumentation amplifier. An
offset voltage can be applied independently of the main supply
through a second INA128, to conduct square and sine
wave tests.
For calorimeter probes, a complementary circuit to use the
lateral wires has also been designed, as shown in figure 3.1 (b). A constant current
source REF200 drives 100 µA through the lateral
wires mounted in series. The wire voltages are measured by two
INA128 instrumentation amplifiers.
(a) CTA control loop. |
(b) Lateral wire measurement circuit. |
Figure
3.1.
Conditioning circuit for SiC
calorimeters.
#3.1.3 Circuit board realization
Photograph of the CTA circuit, plugged into a mounted SiC calorimeter.
The PCB was designed with KiCad.
#3.1.4 Frequency response assessment
Determining the true bandwidth of the hot wire in a closed-loop configuration is quite complex: the ideal way is to apply a direct flow step, i.e. a sudden velocity change. Because it is very difficult to achieve in practice, a workaround has been found to mimic a flow step as closely as possible. Under steady flow, a voltage step (i.e. a square wave) is sent into the feedback loop to create a sudden imbalance. It has been shown by Freymuth (1977)25 that it is equivalent to a flow velocity step, under the assumption that the wire-controller behaves like a third-order system. Although this assumption is not trivial to assess in reality, the CTA has been tested with a SiC calorimeter in a turbulent boundary layer, and a square-wave test has been performed whose results are shown in section 3.2.5.