#3.2 Calibration of the 3C–SiC calorimeter in a turbulent boundary layer
The indirect nature of the thermal skin friction measurement means a calibration curve has to be determined in order to link the sensor output to actual skin friction values. The ONERA Lille operates a turbulent boundary layer wind tunnel whose velocity has been fully characterized using two-axis hot-wire profiling.
#3.2.1 Wind tunnel facility
The ONERA Lille boundary layer wind tunnel facility (figure 3.3) served as the platform for this calibration experiment. It is a zero-pressure-gradient, low-velocity Eiffel-type wind tunnel, with a square test section of 30 cm width. A 2 mm-thick rod followed by a 20 mm-long abrasive strip is installed near the air intake to generate a turbulent boundary layer on the left wall.
Boundary layer wind tunnel facility at ONERA Lille.
#3.2.2 Flush-mounting of sensors on the wind tunnel wall
The SiC sensor is flush-mounted on a wall panel, at x = 1.52 m from the rod, as well as a reference Dantec 55R47 hot-film sensor powered by a StreamLine Pro CTA. The boundary layer can be considered two-dimensional, so we assume that the SiC sensor and hot-film respond to the same wall shear stress values. The sensor die was positioned within a milled pocket of a support PCB, ensuring all surfaces remain flush with the wind tunnel wall (figure 3.4 (b)). Wire bonding provides electrical connection between the die and the PCB, the latter also serving as a mechanical support and a routing medium for electrical contacts. A connector soldered to the backside allowed access to the contacts and enabled a cable connection between the sensor and the main controller board. The support PCB was glued to a plug that fits tightly within the sensor wall panel, ensuring secure mounting (figure 3.4).
(a) Schematic of the resin-printed plug that fits into the wind tunnel wall, allowing for a flush integration of the SiC sensor. |
|
Figure
3.4.
Sensor integration on the wind tunnel
plug.
#3.2.3 Measurement setup
Because of the relatively high voltages needed to power the SiC heater, the CTA output lies in the 40 V range, limiting the choice for adequate digitization instruments. The DEWESoft Sirius acquisition system was used to benefit from its 24-bit resolution at 100 V range, as well as its 200 kHz acquisition rate, ensuring accurate steady and fluctuation measurements. The CTA output voltage, the lateral wire measurements as well as the hot film control circuit’s output are acquired independently and numerically processed (figure 3.5).
Instrumentation setup of the wind tunnel for SiC sensor calibration. The SiC calorimeter is controlled with the in-house CTA and lateral wire measurement electronics, and a Dantec 55R47 hot-film probe is used as a reference measurement.
Thirty-second-long acquisitions were carried out at a sampling frequency of 200 kHz for steady freestream velocities ranging from 0 to 45 m/s in order to calibrate the SiC heater/CTA controller for wall shear stress values. The wind tunnel’s boundary layer was profiled using a motorized hot wire stage for each speed step, and the corresponding wall shear stress was calculated using the Coles-Fernholz method, (Fernholz & Finleyt 1996)131 yielding values up to 2.5 Pa. In addition, the dynamic response of the CTA was assessed by performing square-wave tests under a freestream velocity of 10 m/s, by injecting a 300 mV peak-to-peak square wave voltage to the offset input of the circuit (offset voltage shown in figure 3.1).
#3.2.4 Wall shear stress calibration
#The Coles-Fernholz method
For each speed step, the Coles-Fernholz indirect wall shear stress assessment technique is used. For a x = 1.52 m position of the sensor, the one-dimensional velocity profile u(y) enables calculation of the boundary layer momentum thickness \theta: \theta = \int_0^\delta \frac{u(y)}{U_\infty}\left( 1 - \frac{u(y)}{U_\infty}\right)dy, with \delta the boundary layer thickness. The skin friction coefficient C_\text f is expressed: C_\text f = 2\left( \frac{1}{k} \ln \left(\frac{ \theta U_\infty}{\nu} \right) + C \right)^{-2}\kern -12pt, where k = 0.384 and C = 4.127. Ultimately, the wall shear stress \tau is calculated for each speed step following: \tau = \rho U_\infty^2 C_\text f /2.
The CTA output voltage is time-averaged for thirty seconds and expressed as a variation relative to the no-flow condition (figure 3.6): \text{CTA response} = \frac{E_\text O (\tau) - E_\text O(\tau = 0)}{E_\text O(\tau = 0)}. The plotted error margins show the standard deviation of the fluctuation for each recording, which is mainly due to the frequency content of the turbulence. The resulting calibration curve shows good similarity with measurements performed using the Dantec 55R47, as well as with results reported in the literature for comparable surface-mounted hot-wire wall shear stress sensors. (Ghouila-Houri et al. 2017)132, (Chamard et al. 2023)133
#Calorimetric response to wall shear stress
Calibration curve of the SiC heater operated by the in-house CTA.
Lateral wire differential measurement versus overheat voltage, at constant 30 m/s freestream velocity and no-flow condition.
In order to highlight the unconventional behavior of the SiC’s TCR and its impact on lateral wire measurement, we measured the lateral wire resistance difference in a 30 m/s freestream velocity for different overheat values. We expressed it relative to the baseline resistance difference at 0 m/s (figure 3.7). The latter is non-zero, which is due to the process-induced geometric variability of the wires (figure 2.35). Two distinct regimes appear, which correspond to the lateral wires being respectively in negative and positive PCR ranges. This implies the existence of an overheat value at which the differential lateral measurement is cancelled out.
#3.2.5 Square-wave test
#Principle and assumptions
The square-wave test was introduced and theorized by Freymuth (1967)24, as a way to electronically assess the frequency response of a CTA. Because turbulent flow is a multi-scale and consequently a wideband phenomenon, a frequency spectrum shape does not indicate whether the cutoff comes from the lack of energy in the turbulence itself, or is due to the instrument performance. The natural way to assess this would be to generate a step input in flow speed, i.e. creating an instantaneous speed change. While it is possible using an acoustic tube or a shockwave generation setup, it is a specific experimental procedure that was not available at the time the tests were carried out. A common workaround is to work at a given flow speed and introduce a step perturbation in the feedback loop. This is much more convenient, requiring nothing more than adding a function generator to the overall setup. The CTA compensates this perturbation to bring the heater back to the overheat point, and a timespan t_0 is read between the time of perturbation injection and the time where the CTA first crosses its original value. Freymuth estimates the cutoff frequency of such a system to be f_\text{c} \approx 1/(1.3t_0). However, this estimation works with the assumption that the controller is a 2nd order linear system working in its critically damped regime.
#Assessment of the CTA’s dynamic response
The cutoff frequency of the CTA system is evaluated using a 100 mV step perturbation under a 10 m/s flow, with the resulting transient output shown in figure 3.8. Following Freymuth (1967)24, the first zero crossing occurring at t_0 = 113 µs allows an approximate estimate of the cutoff frequency of our in-house CTA: f_\text{c} \approx 1/(1.3t_0) = 6.8 kHz.
CTA response for a 300 mV peak-to-peak square wave at 10 m/s freestream velocity.
#Turbulence spectra and current limitations of the control electronics
To compare the full dynamic response of the in-house CTA / SiC heater system to that of the commercial hot-film, the energy spectra have been calculated and integrated over 30 seconds, for each velocity step from 0 to 45 m/s (figure 3.9). The energy density is calculated with a frequency-weighted power spectral density fE(f), and is expressed in arbitrary units. The in-house system presents a resonance peak at 4 kHz, which masks out a great part of the frequency information present in the turbulent boundary layer. This highlights the improvements that need to be made on the controller system, to tune its dynamic response to the specific values of SiC resistances, to fully leverage the potential given by the miniaturization work.
Comparison of the energy spectra obtained with the SiC/in-house CTA and Dantec hot-film/Streamline CTA.